Oxygen Transmission Rate (OTR) Analyser


Lyssy PBI Dansensor


Oxygen Permeability Tester OPT 5000



  • This instrument is under NABL scope

Technical Specifications:

  • Measuring range: 0.1-10 cc/m2/day as standard
  • Sensor type: Ceramic solid state oxide sensor
  • Carrier gas: Nitrogen with low oxygen content (eg. N2 5.0/ UHP/ White spot grade)
  • Test gas: Oxygen 1-100% (100% or 20.9% recommended)
  • Sensor life: > 5 years
  • Cascade measuring: Automatic sequential run of multiple test programmes

Measure conditions:

  • Dry measurements: Temperature Range: 5-60 oC
  • Humid measurements: 20-90% RH symmetrically both sides on the following temperature range: 20-40 oC, Limited RH ranges at temperatures 10-20 oC and 40-50 oC

Sample requirements:

  • Measured area: 42 cm2
  • Sample thickness: upto 5mm
  • Minimum sample size: 10*10 cm

Sample to be tested in the OPT-5000 may be placed in sample card which fixes the sample correctly in the chamber which is located behind the lid.

  • The sample must be cut to a size of 10 x 10 cm using the supplied template.
  • Fold the card along the folding line, so it folds easily and place it on the table with the inside facing upwards.
  • Place the sample with the face up above the hole on the short side of the card. The sample must cover the hole; it must not protrude over the edge of the card or the fold. There must be a bit of empty space around the sample.
  • Fold the long side of the card over the sample and press it down over the short side of the card so it is straight. Use a blunt instrument to press along the edges to make the glue stick.





OTR (cc/m2/day)

INR 1500.00 / sample

USD 150.00 / sample